42 · GuardOne
®
· Manual
Activating the automatic life time surveillance
1.
To navigate to the Home page, press Tap to unlock.
2.
Press Settings.
3.
Press Replace UV Light.
4.
Enter the 10-digit serial number shown on the sticker on the
packaging of the UV lamp.
5.
Press Enter to finish the installation procedure.
The GuardOne® Workstation is now ready to be used again.
The remaining life time of the UV bulb is 9000 hours.
Exchanging the battery
The GuardOne® is equipped with a battery permitting device data storage when the device is switched off. Depending on the service life of
the battery, this should be exchanged every 3–5 years.
WARNING
Data loss when removing the batterye
Software version V2.9 and lower: Contact your local Starlab sales partners or Starlab service technician, as a software update to
version V3.0 must first be performed before removing the battery to prevent the loss of the remaining service life of the UV lamp
and the HEPA filter.
Information on the currently installed software version is displayed on the „Standby“ screen (see figure 27).
Additionally, this information is also displayed under „Instrument Info“.
Removing the battery
1.
Switch off the GuardOne® workstation via the mains/power
switch on the rear of the housing.
2.
Open the lid of the upper housing part until the guiding rails to the
left and right engage and keep the lid open.
3.
The battery is located on the rear of the LCD touch screen on the
lid of the upper housing part.
If your GuardOne® has no opening on the rear of the touch
screen, contact your local Starlab sales partner or Starlab
service technician. Contact information can be found on the
status screen by pressing „Starlab Service & Order Information“.
4.
Pressing the spring blade to the left releases the battery for
removal.
Fig. 52: Activate the automatic life time surveillance
Abb. 53: Opening the lid of the upper housing part
Abb. 54: Rear of the LCD touch screen with battery