Fig. 37: UV Timer
Operating the UV timer
›
The timer can only be used before the UV lamp has been switched on.
The UV timer is not mandatory for using UV irradiation.
1.
Switch the UV Status to
(1).
2.
To set the timer, press the hour, minute or second buttons (2).
3.
Select the desired values (3) using the + and - button
.
The desired decontamination time has been set.
4.
To start the UV decontamination, switch the UV status (1) to
.
The UV lamp is switched off automatically after the preset time
has elapsed. If you do not set the timer, UV decontamination will
run indefinitely.
The UV lamp and timer will be switched off automatically as soon
as the front door is being opened. Closing the front door again will
not restart the decontamination procedure.
An acoustic signal will sound as soon as the pre-set time has elapsed.
5.
To exit the page, use Back or Home.