Operation
WARNING!
Incomplete decontamination due to permanent draught!
Due to the permanent draught in the glass chamber, a decontamination of 100 % by irradiation of the UV lamp is not possible.
It is only possible to decontaminate the working environment by 99.995 %.
- Do not work with pathogenic laboratory samples.
- Follow the recommendations in this operating manual.
- For further information about the decontamination of the GuardOne® Workstation, refer to our homepage.
CAUTION!
Health damage due to UV irradiation!
Long-term exposure to UV light results in damage to health.
- For safety reasons, the UV lamp cannot be switched on while the front door is open.
- For safety reasons, the UV lamp will be switched off automatically as soon as the front door is being opened. Closing the door again will
not restart the decontamination procedure.
NOTICE!
Property damage due to incorrect power supply!
An incorrect power supply will lead to property damage.
- Only connect the unit to power sources that match the electrical specifications on the nameplate of the unit.
- Only use sockets with proper electrical grounding.
6.1
Get to Know the Graphical User Interface
The GuardOne® Workstation is equipped with a graphical user interface which is operated via a touchscreen.
Precondition:
›
The main power switch is turned to ON.
Get to know the Standby menu
The Standby page is shown:
›
immediately after switching on the GuardOne® Workstation.
›
after exiting the Home page using the Standby button in the
navigation bar.
The Standby page informs you about the currently installed software
version (1). In addition, if in use, the lab timer (2) will display the remaining
time till the alarm will sound.
Exit the Standby page:
Press Tap to unlock (3) in the navigation bar. You are now on the Home page.
Get to know the Home menu
The Home page offers access to the main functions of the GuardOne®
Workstation.
Reaching the Home page from any other page:
Press Home in the navigation bar. You are now on the Home page.
Reaching the Home page from Standby page:
Press Tap to unlock in the navigation bar. You are now on the Home page.
A
Information bar
B
Display area
C
Navigation bar
1
Date and time
2
Quick Start
3
Airflow & Light
4
UV bulb & Filter Lifetime
5
UV 254 Decon
6
Settings
7
Instrument Info
8
Standby
9
Status of the
GuardOne®
Workstation
10
Lab Timer