Adjusting the Lab timer
The lab timer is working independent of any other function of the GuardOne® Workstation.
You can switch the GuardOne® Workstation into standby mode without affecting the lab timer.
1
To use the lab timer, press the Lab timer symbol in the right corner of
the navigation bar of any page. The Lab timer page opens.
2.
To set the lab timer, press the buttons for hour, minute and second (1).
3.
Enter the desired time values using the + and - button
(2).
4.
Press Start/Pause (3) to start or pause the timer.
After the set time has elapsed, an alarm will be sounded. The alarm
will stop automatically after approx. 5 seconds.
5.
Press End + Reset (4) to stop and reset the timer to 00:00:00.
6.
To exit the page, use Back or Home.
6.2
Operating the Display Area
Quick Start and Quick Stop
When Quick Start is pressed, the GuardOne® Workstation will start working using pre-set parameters for blower speed and LED lights
(factory settings). The factory settings are:
›
Blower speed is 50 %, Laminar Flow Model required.
›
LED light is turned off.
You can change the factory settings at any time during operation by pressing Settings followed by Quick Start Settings.
Activating Quick Start
1.
Open the front door.
In case the front door is closed, the built-in magnetic switch will
prevent the blower from working.
The function of the LED light is independent of the front door.
If the GuardOne® Workstation is in UV decontamination mode,
the UV lamp will be switched off too.
2.
Press Quick Start (1).
The GuardOne® Workstation starts working.
Activating Quick Stop
After Quick Start has been pressed, the button will change to Quick Stop.
Press Quick Stop (1) to switch off the blower and LED light.
The GuardOne® Workstation stops working.
Fig. 29: Adjusting the lab timer
Fig. 30: Quick Start
Fig. 31: Quick Stop