Battery status
Battery State
CADD®-Solis Pump Status
25% to 100%
No alarm
Low battery
• Transition to low battery condition
• Battery low message appears
• Pump emits 3 beeps every 5 min
• Low battery warning message appears on pump display
• Pump is operable
• LCD backlight flashes for 12 ms during each motor operation
Depleted battery • Transition to depleted battery condition
• Battery depleted message appears
• Pump emits a continuous, variable-tone alarm
• Depleted battery warning message appears on pump display
• Battery power is too low to operate pump
• Pump delivery operation stops
Shut down
Pump shuts off due to too low operating voltage.
Alkaline battery life with screen backlight intensity set to 3
These estimates are based on laboratory tests conducted at room temperature using new batteries.
Actual battery life varies depending on the battery brand, shelf life, temperature conditions, delivery
rate, and frequency of screen display and backlighting. It is recommended that new batteries be kept
available for replacement.
PCA and Continuous delivery
Delivery Rate
(mL/hr)
Operating Time
(hrs)
Vol. Delivered
(mL)
0.4
260
104
1
.0
210
210
5
.0
165
825
10
.0
150
1500
30
.0
70
2100
50
.0
42
2100
100
.0
30
3000
200
.0
22
4400
350
.0
15
5250
500
.0
7
3500
Intermittent delivery
Volume
(mL)
Duration
(hr)
Cycle
(hr)
KVO
(mL/hr)
Operating Time
(hrs)
Vol. Delivered
(mL)
20.2
1
4
0.2
85
430
40.7
.0
1
4
0.2
72
733
61
.0
1
6
0.2
91
925
125
.0
1
6
0.2
70
1460
200
0
1
12
0.2
52
1300