107
Tasks and Advanced Tasks
Ta
sk
s a
nd
A
dv
anc
ed T
as
ks
Reservoir Empty Alarm Type
The reservoir empty alarm type option allows you to select between two types of alarms when the reservoir
is empty:
One Time Only:
When a reservoir empty alarm occurs, you must press a button to acknowledge or silence
the alarm. This alarm does not reoccur once it has been acknowledged.
Repeat Until Removed or Reset:
When a reservoir empty alarm occurs, you must press a button to
acknowledge or silence the alarm. The alarm is repeated every 5 minutes until the reservoir volume has
been set to a new value, the reservoir is removed, or the pump is powered down.
To set the reservoir empty alarm type:
1. Make certain the pump is stopped.
2. In the Alarm Settings menu, press
U
or
D
to highlight
Reservoir
Empty Alarm Type
and press
S
.
3. Unlock the keypad.
4. Press
U
or
D
to highlight
One Time Only
or
Repeat Until
Removed or Reset,
and select
Save.
PM Reminder On/Off
If your facility establishes a maintenance program for the pump, you can use the PM reminder on/off
option to program the pump to sound an alarm when it is time for preventive maintenance. The message
begins appearing on the date programmed and during every power up until the date is reset. Use the PM
reminder on/off screen to turn the alarm on and then continue to the PM reminder screen to specify the
interval at which the message should appear, or use it to reset the reminder.
To turn the PM reminder on or off:
1. Make certain the pump is stopped.
2. In the Alarm Settings menu, press
U
or
D
to highlight
PM
Reminder On/Off
and press
S
.
3. Unlock the keypad.
4. Press
U
or
D
to highlight
Off
or
On,
and select
Save.
5. If you set the
PM Reminder On/Off
to On, press
U
or
D
to highlight
PM Reminder
and press
S
.
Instructions for scheduling the PM Reminder are in the next section.