130
References and Troubleshooting
Re
fe
re
nc
es a
nd
Tr
ou
bl
es
ho
ot
ing
Bolus accuracy specification: ± 6%
Actual test data for bolus accuracy at 0.05 mL:
Average
0.0508 mL
% Error
1.6%
Minimum Error %
–3.0%
Maximum Error %
4.2%
Actual test data for bolus accuracy at 20 mL:
Average
20.108 mL
% Error
0.5%
Minimum Error %
–2.9%
Maximum Error %
4.5%
Maximum volume infused under single-fault conditions
CADD
®
administration set: 0.15 mL
CADD
®
high volume administration set: 0.30 mL
Delivery rate during priming
Approx. 250 mL/hr with standard volume cassette tubing
500 mL/hr with high flow volume cassette tubing
Alarm disabled during priming
Air-In Line
Delivery Specifications
Common Delivery Specifications
Reservoir volume
0 to 9999
Programmable in 1 mL increments
Displayed in 0.1 mL increments
Given
0 to 99,999.99 in 0.01 unit increments
Res vol low trip point
1 to 999 mL in increments of 1 mL
Res vol empty alarm
•
Insistent and one time only
•
Non-insistent and repeating
Delayed start
0 to 96 hr in 5 min increments
Pump stopped alarm
Informational or high priority
Air detector
On or off
Air detector sensitivity
Low sensitivity: Single bubble > 400 μL
High sensitivity: Single bubble > 150 μL
Alarm volume
High, medium, or low