35
Pump Programming
Pu
mp
Pr
og
ra
mmi
ng
Editing Delivery Settings
To view and edit delivery settings:
1. Stop the pump if it is running.
2. In the Tasks menu (see page 88), press
U
or
D
until
View Delivery
Settings
is highlighted, and then press
S
.
3. Press
U
or
D
until the desired setting is highlighted, and then
press
S
.
4. If requested, unlock the keypad.
5. Press
U
or
D
until the desired value appears on the screen, then
select
Save.
NOTE:
If the new value is above the
maximum soft limit or below the
minimum soft limit, it appears in
amber on the screen. To select a value
above the maximum soft limit or
below the minimum soft limit, select
Confirm
and then select
Yes.
Change any additional settings by scrolling through the remaining delivery settings until all values are
correct.
NOTE:
Changes to any of the delivery settings must be confirmed and accepted in a pump settings
review. The confirmation requires a security code and must take place before the pump will start.
Press stop/start
9