Specifications (Nominal)
General Pump Specifications
Used to test the pump:
•
CADD
™
medication cassette reservoirs,
<
21-7002
•
CADD
®
extension sets,
<
21-7045
•
CADD
®
administration sets,
<
21-7091
•
CADD
®
high volume administration sets,
<
21-7055
Resolution
•
CADD
™
medication cassette reservoir: 0.050 mL per pump stroke nominal
•
CADD
®
administration set: 0.050 mL per pump stroke nominal
•
CADD
®
high volume administration set: 0.1 mL per pump stroke nominal
Size
4.1 cm × 10.2 cm × 12.7 cm (1.6 in × 4 in × 5 in), excluding cassette and accessories
Weight
595 g (21 oz) including 4 AA alkaline batteries, excluding other accessories
Pump alarms
High priority alarms:
•
Air in line detected
•
Battery depleted while delivering
•
Battery removed while delivering
•
Battery unusable while delivering
•
Disposable attached improperly
•
Disposable detached while delivering
•
Disposable locked but not latched
•
Disposable type high flow administration set not allowed
•
Disposable type high flow administration set required
•
Disposable type invalid
•
Downstream occlusion
•
Key stuck
•
Pressure sensor faulty
•
Pump automatically stopped
•
Rechargeable battery end of life
•
Remote dose cord key stuck
•
Reservoir volume empty
•
Stop mode reminder
•
Upstream occlusion
Medium priority alarms:
23
Low priority alarms:
10
Informational messages/alerts:
23
Battery fallout alarm
Alarm sounds for 2 minutes if the pump has been powered up for a minimum of 2 minutes.
NOTE:
Alarm enabled while pump is in run mode only.