Power sources
• AC adapter
• CADD®-Solis rechargeable battery pack
• Four AA (IEC LR6) alkaline batteries
Charging system for internal memory backup battery
The internal memory backup battery uses lithium manganese dioxide technology. It charges whenever
the pump is powered on and has a 10-month memory capacity once it has been charged for 250 hours
at 20°C (68°F).
System operating temperature
2°C to 40°C (36°F to 104°F)
System storage and transportation temperature
–20°C to 60°C (–4°F to 140°F)
Relative humidity
20% to 90% relative humidity, non-condensing
Atmospheric pressure
70 kPa to 106 kPa (10.2 psi to 15.4 psi)
System delivery accuracy
WARNING:
•
Ensure that the ± 6% system delivery accuracy specification is taken into account when
programming the pump and/or filling the reservoir. Failure to do so may result in medication in
the reservoir becoming depleted sooner than expected. If the pump is being used to deliver critical
or life sustaining medication, the interruption in the delivery of medication could result in patient
injury or death.
•
System delivery inaccuracies may occur as a result of back pressure or fluid resistance, which
depends upon drug viscosity, catheter size, and extension set tubing (for example, microbore
tubing), and placing the infusion reservoir and/or pump above or below the level of the patient.
System delivery inaccuracy may result in under or overdelivery of medication.
± 6% (nominal). At low infusion rates, this accuracy may not be achieved for short periods. During the
total infusion time, the accuracy averages out.
System definition
CADD
®
-Solis pump with:
•
An attached medication cassette reservoir and CADD
®
extension set,
or
◆ An attached medication cassette reservoir with Flow Stop feature and CADD
®
extension set,
or
•
An attached CADD
®
administration set
or
◆ An attached CADD
®
administration set with Flow Stop feature
High pressure alarm threshold
18 ± 9 psi (1.24 bar ± 0.62 bar)
Air detector alarm
Sensitivity:
•
Low: Single bubble > 400 μL
•
High: Single bubble > 150 μL
Accumulated Air:
Greater than 1 mL air over 15 minutes (nominal)