92
Tasks and Advanced Tasks
Ta
sk
s a
nd
A
dv
anc
ed T
as
ks
Time and Date
The Change Time and Date setting allows you to edit the time and date, and to choose a time and date
format. The time and date options should reflect the current time and date. The clock is powered by a
separate, internal battery, which retains the time and date even when the 4 AA batteries or the battery
pack is removed. The pump uses this feature to record the time and date of events in the delivery and event
logs, and in other reports. It is also used to determine when delayed starts and next dose start times begin,
and when preventive maintenance alarms occur.
NOTE:
The internal battery is rechargeable and automatically recharges when the pump is powered on. If
the pump is off for a long period of time, it may not show the correct time and date when it is powered
on. Check the time and date, and edit these settings if necessary.
CAUTION:
If delivery of an infusion is affected by a time or date change, an alarm message appears
and must be confirmed.
In the Tasks menu, press
U
or
D
to highlight
Change Time and
Date
and press
S
.
Current Time
To set the current time:
1. In the Change Time and Date menu, press
U
or
D
to highlight
Current Time
and press
S
.
2. Press
U
or
D
to scroll to the correct hour and press
S
to
navigate to the minutes.
3. Press
U
or
D
to scroll to the correct minutes and press
S
to
navigate to the am or pm field.
4. Press
U
or
D
to scroll between am and pm and press
S
when
the correct setting appears. If you are using a 24-hour clock, the am/pm
field does not appear. Select
Save.