30
Pump Programming
Pu
mp
Pr
og
ra
mmi
ng
To start a new patient and manually program the protocol:
1. Make sure the pump is stopped.
2. In the Advanced Tasks menu (see page 99), press
U
or
D
to highlight
Start New Patient
and press
S
.
3. Unlock the keypad.
4. Press
U
or
D
to highlight a therapy and press
S
.
5. Press
U
or
D
to highlight
[Program
Manually]
and press
S
.
If you are programming a PCA therapy,
you must also press
U
or
D
to
highlight the desired units and press
S
. If the units are mg or mcg, press
U
or
D
to highlight the desired
concentration and select
Save.
6. A screen appears with a summary of your selections. Confirm that the
correct information appears, and select
Yes.
NOTE:
If the information is incorrect, select the left soft key as many
times as required to back out of the screen, allowing you to start
over.
7. The “Review pump settings” screen appears. Select
Review.
8. Press
U
or
D
to scroll through all the settings. If the highlighted
setting is correct, select
Accept Value.
A green check mark appears
next to the setting.
If the setting is incorrect, press
S
to edit. Press
U
or
D
to
modify the setting and then select
Save
or
Confirm
(see note).
NOTE:
If you edit any of the parameters, you may exceed the soft limit
range and must
Confirm
the soft limit override on each screen. The
parameter and value will be displayed in amber.
WARNING:
The manual program mode does not contain programming limits. Carefully review
each parameter to ensure it accurately matches the prescription. Failure to set the manual
program to the correct values could result in serious patient injury or death.
9. When the review is complete, select
Next.
10. Follow the instructions on the pump to attach the cassette, prime, and
start the pump.