Taper Delivery Mode
Ta
pe
r
Pr
og
ra
mmi
ng
Programming Example
WARNING:
Ensure that the ± 6% system delivery accuracy specification is taken into account when
programming the pump and/or filling the reservoir. Failure to do so may result in medication in
the reservoir becoming depleted sooner than expected. If the pump is used to deliver critical or life-
sustaining medication, the interruption in the delivery of medication could result in patient injury
or death.
Example:
A total of 1 L of TPN solution must be delivered to the patient over 12 hours. The TPN bag
contains 1050 mL of solution. Delivery must taper up over 90 minutes at the beginning of the delivery, and
taper down over 90 minutes at the end of delivery.
1. Select the Therapy
• Stop the pump if it is running.
• Select
Tasks,
then
View Advanced Tasks,
and then
Start New
Patient.
• Unlock the keypad.
• Press
U
or
D
to highlight the
Taper
therapy and press
S
. See
on page 27 if you need more information on selecting
the therapy.
2. Select the Qualifier
• Press
U
or
D
to highlight the desired qualifier and press
S
.
3. Select the Drug
• Press
U
or
D
to highlight the desired drug, and press
S
.
4. Confirm and Review the Settings
• Review the information on the screen and select
Yes
if they are
correct, or
No
to edit the settings.