Step Delivery Mode
St
ep
Pr
og
ra
mmi
ng
Step Down
This option allows you to reduce the rate of infusion at any time. You cannot step down lower than the
initial rate. Once the step down task has been performed successfully, the automatic step increases after
each step duration are cancelled. Delivery continues at the step down (hold) rate. The step up task can be
used to restart the automatic step increases.
To step down from the current rate:
1. In the Advanced Tasks menu, press
U
or
D
to highlight
Step Down
and press
S
.
2. The pump displays a screen that tells you the current rate, and asks you
if you want to step down to the previous step rate. Select
Yes.