General Description
G
ene
ra
l
D
es
cr
ip
ti
on
3. Depending on the level of security
required and how the CADD
®
-Solis
system administrator has programmed the
security settings, you may be able to use
the pump key to unlock the keypad. When
accessing a menu or function that requires
a security code, 1 of 2 prompts appears, as
shown in the sample screens to the right.
To unlock the keypad with the pump key, turn the key counterclockwise.
NOTE:
It is possible for the keypad to be locked while the cassette/keypad
lock is unlocked. To use the key to unlock the keypad, first lock the
cassette/keypad lock, and then unlock it.
Customized Security Settings
The keypad code, clinician code, and administrator code may be customized by the CADD
®
-Solis
system administrator when setting up protocols with the CADD
®
-Solis Medication Safety Software, or
in the security settings under Advanced Tasks. See your CADD
®
-Solis system administrator for more
information, or to learn the security code you should use if the codes have been customized.
The security settings option in the Advanced Tasks menu allows you to customize the pump security
codes. The keypad code, clinician code, and administrator code can all be customized to a unique three-
digit number from 001 to 899. If desired, all the codes may be identical. In that case however, all users
would have the highest permission level assigned to that code. For example, if the keypad code and
clinician codes are the same, each time the code is entered, the user has clinician code permissions. See the
security level tables beginning on page 19 for the code permissions.
To access Security Settings:
1. From the home screen, select
Tasks.
2. Press
U
or
D
to highlight
View Advanced Tasks
and press
S
.
3. Press
U
or
D
to highlight
Security Settings
and press
S
.
Either the key or a code
unlocks the keypad.
Only a code unlocks the keypad.