41
PCA Delivery Mode
PC
A
Pr
og
ra
mmi
ng
• Press
U
or
D
until the value reads
5 mg/hr
and select
Save.
• The pump returns to the Review screen.
6. Enter the PCA Dose
• Press
U
or
D
to highlight
PCA Dose.
If the value shown is set at
the desired dose, select
Accept Value
and go to step 7. If the value
shown does not show the desired dose, press
S
.
• Press
U
or
D
until the value reads
2 mg
and select
Save.
• The pump then returns to the Review screen.
7. Enter the PCA Lockout
• Press
U
or
D
to highlight
PCA Lockout.
If the time shown is set
at the desired value, select
Accept Value
and go to step 8. If the value
shown does not show the desired time, press
S
.
• Press
U
or
D
until the value reads
15 Min
and select
Save.
• The pump then returns to the Review screen.