128
References and Troubleshooting
Re
fe
re
nc
es a
nd
Tr
ou
bl
es
ho
ot
ing
Step delivery
Volume to
Deliver
(mL)
Starting
Rate
(mL/hr)
Step
Duration
(min)
Step Rate
Increase
(mL/hr)
Max
Rate
(mL/hr)
No. of
Steps
Operating
Time
(hrs)
900
45
15
45
315
7
17
1500
37.5
30
80
300
5
14
2500
30
30
90
300
4
14
Taper delivery
Volume
(mL)
Period
(hr)
Taper Up
(hr)
Taper Down
(hr)
KVO
(mL/hr)
Operating Time
(hrs)
Vol. Delivered
(mL)
2000
10
1
1
5
20
3360
3000
10
1
1
5
16
3950
Classification
CF
J
, Class II
K
Moisture protection
Splashproof (
E
) per IEC 60529
Maximum infusion pressure
1.86 bar (27.0 psi)
Maximum time to occlusion alarm and Bolus volume at occlusion alarm
Flow Rate
(mL/hr)
Tubing Set
Maximum Time to Occlusion
Bolus at Occlusion
Raw Test Data
(minutes)
Specification
(minutes)
Raw Test Data
(mL)
Specification
(mL)
0.1
CADD™ medication
cassette reservoir
<
21-7002 with CADD
®
extension set
<
21-7045
90
≤ 160
0.107
≤ 0.25
CADD
®
administration set
<
21-7091
122
≤ 190
0.139
≤ 0.30
CADD
®
high volume
administration set
<
21-7055
1140
≤ 1200
1.250
≤ 1.40
Flow Rate
(mL/hr)
Tubing Set
Maximum Time to Occlusion
Bolus at Occlusion
Raw Test Data
(seconds)
Specification
(seconds)
Raw Test Data
(mL)
Specification
(mL)
150
CADD™ medication
cassette reservoir
<
21-7002 with CADD
®
extension set
<
21-7045
4
≤ 45
0.069
≤ 0.25
CADD
®
administration set
<
21-7091
4
≤ 45
0.072
≤ 0.30
CADD
®
high volume
administration set
<
21-7055
32
≤ 90
1.059
≤ 1.40