Patient Permissions
Security levels are used to limit unauthorized access to certain programming and operating functions.
In some cases, the default security settings may be changed to allow patients additional access to pump
functions. Do not reduce security settings without providing proper training to patients.
WARNING:
•
Do not disclose the pump security codes or any other information that would allow the patient
or unauthorized clinician complete access to programming and operating functions. Improper
programming could result in serious patient injury or death.
•
Do not leave the pump unattended while unlocked. All programming functions are accessible while
the pump is unlocked, and improper programming could result in serious patient injury or death.
In the Advanced Tasks menu, press
U
or
D
to highlight
Patient
Permissions
and press
S
.
Priming Security On/Off
Setting this value to Off allows patients to prime the tubing without having to enter a security code.
To set patient permissions for priming activity:
1. Stop the pump if it is running.
2. In the Patient Permissions menu, press
U
or
D
to highlight
Priming Security On/Off
and press
S
.
3. Unlock the keypad.
4. Press
U
or
D
to set the security to
On
(security code required) or
Off
(no security code required) and select
Save.