27
Pump Programming
Pu
mp
Pr
og
ra
mmi
ng
Protocol Selection from a Library
Start New Patient
Each time a new patient is started, it is recorded in the event log. All other reports are cleared. Protocol
libraries are created with the CADD
®
-Solis Medication Safety Software and are loaded onto the pump by the
CADD
®
-Solis system administrator. You may select a standard therapy protocol from a library created by
the CADD
®
-Solis administrator, or manually program the pump.
Start New Protocol Same Patient
The process for starting a new protocol for the same patient is the same as starting a new patient. However,
the event log does not insert a new patient marker. All other reports are cleared except the delivery log.
NOTE
: When starting a new protocol, attach a
new
reservoir with the proper medication.
To start a new patient and run an existing protocol from a library:
1. Make sure the pump is stopped.
2. In the Advanced Tasks menu (see page 99),
press
U
or
D
to highlight
Start New
Patient
or
Start New Protocol Same
Patient
and press
S
.
3. Unlock the keypad.
4. Press
U
or
D
to highlight a therapy and press
S
.
5. Press
U
or
D
to highlight a qualifier and press
S
.
6. Press
U
or
D
to highlight a drug and press
S
.
7. A screen appears with a summary of your selections. Review them to
confirm that you have entered the correct therapy, qualifier, and drug.
Select
Yes.