Step delivery
Volume to
Deliver
(mL)
Starting
Rate
(mL/hr)
Step
Duration
(min)
Step Rate
Increase
(mL/hr)
Max
Rate
(mL/hr)
No. of
Steps
Operating
Time
(hrs)
900
45
15
45
315
7
21
1500
37.5
30
80
300
5
20
2500
30
30
90
300
4
18
Taper delivery
Volume
(mL)
Period
(hr)
Taper Up
(hr)
Taper Down
(hr)
KVO
(mL/hr)
Operating Time
(hrs)
Vol. Delivered
(mL)
2000
12
1
1
5
29
4800
3000
12
1
1
5
17
4220
Rechargeable battery pack life with screen backlight intensity set to 3
These estimates are based on laboratory tests conducted at room temperature using a new CADD
®
-Solis
rechargeable battery pack. Actual battery life varies depending on temperature conditions, delivery
rate, and frequency of screen display and backlighting. It is recommended that new batteries be kept
available for replacement.
PCA and Continuous delivery
Rate (mL/hr)
Life (hrs)
Volume (mL)
0.4
119
47.6
1
.0
89
89
.0
5
.0
88
440
.0
10
.0
86
860
.0
30
.0
43
1290
.0
50
.0
30
1500
100
.0
19
2375
200
.0
16
3200
350
.0
11
3850
500
.0
8
8000
Intermittent delivery
Volume
(mL)
Duration
(hr)
Cycle
(hr)
KVO
(mL/hr)
Operating Time
(hrs)
Vol. Delivered
(mL)
20.2
1
4
0.2
54
273
40.7
.0
1
4
0.2
51
520
61
.0
1
6
0.2
54
550
125
.0
1
6
0.2
43
896
200
0
1
12
0.2
48
800