76
Taper Delivery Mode
Ta
pe
r
Pr
og
ra
mmi
ng
12. Accept Values
• If any values have not already been accepted, press
U
or
D
to
highlight each setting and select
Accept Value
if the setting is correct.
A green check mark appears next to each accepted value.
• After your review is complete, select
Next.
13. Prepare the Pump for the Patient
• Follow the instructions for attaching a cassette, priming, and attaching the pump to the patient in
Starting Daily Infusion
When a new cassette is attached to the pump at the beginning of infusion, the reservoir volume should be
reset.
This also resets the infusion.
Follow the instructions in
beginning on page 78), for
attaching a cassette and resetting the reservoir. Then when the pump is started, delivery will start at the
beginning of the infusion.
Infusion Duration
Plateau Rate
Taper
Up
Taper
Down
Infusion Volume
Time
Deliv
er
y R
at
e
KVO
New cassette attached.
Reservoir volume and infusion profile reset.