Intermittent Delivery Mode
In
te
rmi
tt
en
t
Pr
og
ra
mmi
ng
Stopping the Pump During an Infusion
Stopping the pump
between doses
does not affect the start time of subsequent doses.
Stopping the pump
while a dose is in progress
shifts all subsequent doses by the amount of time the pump
is stopped (see illustration).
If you want to make up for the time lost in the cycle because the pump was stopped during a dose,
wait
until the current dose is completed.
Then, stop the pump and reprogram the Next Dose Start Time. For
additional information, see
Resuming the Dose
To resume delivery of a dose stopped in progress, simply restart the pump by pressing stop/start
9
.
Resetting the Cycle
To reset the cycle, stop the pump between doses, and reprogram the Next Dose Start Time.
When a dose is stopped in progress, the “Interrupted” screen appears. To cancel the current dose and reset
the cycle, reprogram the Next Dose Start Time through the Tasks menu. The next cycle will start at the
time you select.