Continuous Delivery Mode
Con
ti
nuo
us
Pr
og
ra
mmi
ng
6. Enter the KVO rate
• Press
U
or
D
to highlight
KVO Rate.
If the value shown is set
at the desired rate, select
Accept Value
and go to step 7. If the value
shown does not show the desired rate, press
S
.
• Press
U
or
D
until the value reads
0 mL/hr
and select
Save.
The
pump then returns to the Review screen.
7. Enter the Reservoir Volume
• Press
U
or
D
to highlight
Reservoir Vol.
If the value shown
is correct, select
Accept Value
and go to step 8. If the value is not
correct, press
S
.
• Press
U
or
D
until the value reads
1000 mL
and select
Save.
• The pump then returns to the Review screen.
8. Accept Values
• If any values have not already been accepted, press
U
or
D
to
highlight each setting and select
Accept Value
if the setting is correct.
A green check mark appears next to each accepted value.
• After your review is complete, select
Next.
9. Prepare the Pump for the Patient
• Follow the instructions for attaching a cassette, priming, and attaching the pump to the patient in