Electromagnetic Emissions and Immunity Declarations
Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The CADD®-Solis pump is intended for use in the electromagnetic environment specified below. The customer or the user of
the CADD®-Solis pump should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
RF emissions
CISPR 11
Group 1
The Pump uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Pump is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Not applicable
Compliance using:
1. 100-240 VAC 50/60Hz to 7VDC switching desktop AC adapter, with an AC power cord length of 1.8 m
(6 feet).
2. Rechargeable battery pack.
3. Remote dose cord with a length of 152 cm ± 5 cm (60 in. ± 2 in.).
4. USB cable length of less than 2 m (6.5 feet).
WARNING:
• The use of power supplies and a remote dose cord other than those listed in the electromagnetic
emissions declaration may result in increased emissions or decreased immunity of the pump.
•
The pump should not be used adjacent to or stacked with other equipment. If adjacent or stacked
use is necessary, verify normal operation of the pump in the configuration in which it is to be used.
Guidance and Manufacturer’s Declaration —Electromagnetic Immunity
The CADD®-Solis pump is intended for use in the electromagnetic environment specified below. The customer or user of the
CADD®-Solis pump should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment—Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
± 15 kV air
(IEC 60601-2-24)
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5%
U
t (> 95% dip in
U
t) for 0.5 cycle
40%
U
t (60% dip in
U
t)
for 5 cycles
70%
U
t (30% dip in
U
t)
for 25 cycles
< 5%
U
t (> 95% dip in
U
t) for 5 sec
< 5%
U
t (> 95% dip in
U
t) for 0.5 cycle
40%
U
t (60% dip in
U
t)
for 5 cycles
70%
U
t (30% dip in
U
t)
for 25 cycles
< 5%
U
t (> 95 % dip in
U
t) for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Pump requires continued
operation during power mains interruptions,
it is recommended that the Pump be powered
from an uninterruptible power supply or a
battery.
Power frequency
50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
400 A/m
(IEC 60601-2-24)
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE:
U
t is the a.c. mains voltage prior to application of the test level.