55
Intermittent Delivery Mode
In
te
rmi
tte
nt
Pr
og
ra
mmi
ng
6. Enter the Dose Duration
• Press
U
or
D
to highlight
Dose Duration.
If the value shown is
set at the desired duration, select
Accept Value
and go to step 7. If the
value shown does not show the desired duration, press
S
.
• Press
U
or
D
until the value reads
1 Hrs 0 Min
and select
Save.
The pump then returns to the Review screen.
7. Enter the Dose Cycle
• Press
U
or
D
to highlight
Dose Cycle.
If the value shown is set
at the desired cycle, select
Accept Value
and go to step 8. If the value
shown does not show the desired cycle, press
S
.
• Press
U
or
D
until the value reads
6 Hrs 0 Min
and select
Save.
• The pump then returns to the Review screen.
8. Enter the Dose Rate
• In this example, Dose Rate is a calculated value and cannot be edited.