M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 2 of 4
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
15 - 2
Checking the swivel-arm system
4. Set the projection angle to 60 degrees with presetting knob (2/ Fig. 1).
The preset angle is displayed for approx 3 s.
5. Press one of the buttons (1/ Fig. 1).
The rotary motion of the swivel-arm system shall stop automatically when the
preset projection angle (both + and -) is reached and at 0 degrees.
Vertical adjustment
15
1. Press the buttons for upward and downward movement (3/ Fig. 1) of the swivel-
arm system.
The movement shall stop automatically when either end position is reached or
when the button is released.
Fig. 2 Display at the lower part of the stand
All eight rotary motion buttons (four on either side) must be
checked. After the button has been released, the swivel-arm sys-
tem shall remain in the position set.
Remove the protective strips before performing vertical adjust-
ment of the swivel arm.
If not removed, the protective strips could be damaged.
Make sure the protective strips are removed before adjustments.,
see page 3 - 5.
All eight vertical adjustment buttons (four on either side) must be
checked. The switching off in upper and lower end position is
effected by switches S881 and S884 respectively.