Calibrating and adjusting the AEC
11 - 13
Siemens AG
Register 3
SPB7-230.033.11
Page 13 of 22
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
5. Type the ID number of the new correction table (see Table 1 on Page 11-1) in the
dialog which appears. Press <F2> to install the table. A message box will appear
before the correction file has been found.
6. When the transaction of tables is finished, a dialog like Fig. 4, Page 11 - 15 will
appear. To complete the calibration, proceed from step 5 of that section.
In order to resume calibrating after having quit the
program:
Follow the instructions from step number 1 under
"Calibrate correction tables" on Page 11 - 14.