M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 8 of 10
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
10
-
8
Checks
with
high
voltage
Fig.
9
Fig.
10
30 kV
1 V/T (1 T <=> 5 kV)
30 kV
1 V/T (1 T <=> 5 kV)
28 mA
0.5 V/T (1 T = 20 mA)
23 mA
0.5 V/T (1 T <=> 20 mA)
360
ms
50
ms/T
430
ms
100
ms/T
Small focus is selected by attaching a magnification table.
Record the measured kV and mAs value in the test protocol, page
22 - 10.
Tube P40 MoW,
MAMMOMAT 3000 Nova
* tungsten anode 0.15
Tube P40 MoW,
MAMMOMAT 1000/3000 Nova
molybdenum anode 0.15
M
A
M
0
0
1
8
2
30 kV 10 mAs (P = 0.85 kW)
M
A
M
0
0
1
1
5
30 kV 10 mAs (P = 0.70 kW) >>
NOTE
NOTE