Prerequisites
1
-
3
Siemens AG
Register 3
SPB7-230.033.11
Page 3 of 6
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Unless otherwise stated, these instructions describe the stand with pivoted object-table
arm and generator with integrated radiation shield and control panel.
Depending on the installation configuration, some points may be omitted.
Meters and appliances, tools
1
Meters and appliances required
1
•
Power Line Impedance Meter.
•
Protective ground-wire tester.
•
Oscilloscope, e.g TEKTRONIX 314.
•
Digital multimeter (Fluke type 8060A, part no. 97 02 101 Y4290 or Fluke type 87, part no.
97 03 976 Y4290).
•
Sensitometer.
•
Densitometer e.g PDA 81.
•
Service PC with connecting cable (PC-Generator), part no. 99 00 440 RE999.
•
Normi 7 test body (provided by the customer) or SIB phantom Type 42 001 (PTW or
INAK).
•
Centering cross, part no. 96 60 051 RE 999.
•
AEC calibration plexi:
- Three plates, "2" cm (part no. 65 61 232)
- One plate, "1" cm (part no. 65 61 224)
Tools required
1
•
Standard installation tool kit.
•
Torque wrench for bolting the stand/generator to the floor (optional).
•
Electrical screwdriver with adjustable torque is recommended.
•
Calculator
Calibrated instruments are required.
The existing ground conductor in the mains cable must under no
circumstances be disconnected when operating the oscilloscope.
Life-threatening electric shock hazard exist.
For those measurements, in which any resulting ground loop may
falsify the measuring result, use the differential amplifier (differ-
ence measurement).
This symbol indicates exposure of X-ray radiation.
NOTE
WARNING