18 - 1
Siemens AG
Register 3
SPB7-230.033.11
Page 1 of 2
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Checking the exposure blocking
18
DLF switch on
18
Activate the DLF switch. In <Main menu> select <Configuration>, <Miscellaneous> and
<DLF switch>, press space to set switch to ON. Press <F2> to save.
Reduce grid speed
18
Reduce the grid speed. In <Main menu> select <Configuration> and <Grid speed>.
Set the following values:
Cassette loaded switch on
18
Set the cassette loaded switch on.
The "cassette loaded" switch must be set to "ON" with the service
PC program.
The film cassette must be exchanged after each exposure. A sec-
ond exposure before exchanging the cassette is not possible
(exposure release is blocked).
Grid
1 Help 2 Save 3
4
5
6
7
8
9
10
Quit
Grid fast speed time
500
Grid fast speed
80
Grid slow speed
20
ms (2.5 s max)
% of max
% of max
<ESC> to exit, <TAB> move to next entry field
M
A
M
0
0
7
6
2
NOTE
NOTE