M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 14 of 22
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
11 - 14
Calibrating and adjusting the AEC
Calibrate correction tables
11
The Anode/Filter/Speed combinations should be calibrated one at a time.
1. In <Mainmenu> select <Configuration>, <AEC> and <AEC correction tables>.
2. Place the marker by one of the desired Anode/Filter/Speed combinations. Press
<F3> to transfer the table.
3. This message box is displayed while the service PC receives the correction table.
Make sure the backup floppy is inserted.
The dose monitoring function is only enabled if the applicable
combination is calibrated.
All object table groups and Anode/Filter combinations available
at the site must be calibrated prior to release for clinical use.
However, calibration does not have to be attempted for W/Rh
combination in stereo mode.
NOTE
1 Help 2 Inst
Cal
3
4
5
6
7
8
9
10
Quit
M
A
M
0
0
7
3
0
<ESC> to exit, <F2> to install new table, <F3> to calibrate existing
Anode/Filter
Speed
ID
Calibrated
Mag
Grid
No grid
Stereo
Mo/Mo
H
1104
No
No
No
No
Mo/Mo
D
----
---
---
---
---
Mo/Rh
H
1105
No
No
No
No
Mo/Rh
D
----
---
---
---
---
W/Rh
H
1106
No
No
No
No
W/Rh
D
----
---
---
---
---
Existing tables in the AEC:
NOTICE
Receiving correction table
for Mo/Mo H
63% done
M
A
M
0
0
6
0
2
AEC co rrection tables