Calibrating and adjusting the AEC
11 - 19
Siemens AG
Register 3
SPB7-230.033.11
Page 19 of 22
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Recalibration of an object table
11
An already calibrated object group can be recalibrated.
1. Repeat the procedure described in steps 1 to 4, page 11 - 14.
2. To recalibrate a correction table for an object table group, place the cursor in front
of the group. Press <F4>. This will reset the existing calibration of the chosen
object table group and replace it with the original correction table stored on the
floppy.
3. Press <F2> to save.
4. Recalibrate the desired object table group according to the normal procedure,
from step 2, page 11 - 14.
Make sure the backup floppy is inserted.
NOTE