20 - 1
Siemens AG
Register 3
SPB7-230.033.11
Page 1 of 10
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Final procedures
20
Service PC and measuring instruments
20
Saving the configuration file
20
Main menu:
1. Select <Configuration> <Save config file>.
2. Press <F2> to save data.
Reading the exposure counter
20
Main menu:
1. Select <Service> <Show exposure counter>.
2. Enter the current counter reading in the certificate.
Deleting the error memory
20
Main menu:
1. Select <Service> <Delete error buffer>.
2. Delete the error memory with <Y>.
Saving the programmed values
20
Main menu:
1. Select <Backup> <Copy installation area to floppy> <All>.
2. Leave the program with <F10> when the backup is ready.
Make sure that the backup floppy is inserted.
Keep the backup floppy with other documentation for the
MAMMOMAT in question.
NOTE
NOTE