M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 8 of 22
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
11 - 8
Calibrating and adjusting the AEC
*Applicable on condition that the detector is placed in chest-wall position.
Minimum projecting
Length: 10 mm
AEC calibration plexi
Object table
25 mm
Measuring
point O.D.
Film
20 mm
approx
approx
Fig. 2 Position of AEC calibration plexi and film
Metal plate
part no. 65 52 728
Sensitometer strip