Calibrating and adjusting the AEC
11 - 15
Siemens AG
Register 3
SPB7-230.033.11
Page 15 of 22
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
4. When the transfer is done, the dialog
Calibration AEC correction table
appears.
5. To calibrate the correction table for an object table group, place the cursor in front
of the group. Press <F3>.
6. Follow the instructions in the message box above and press <ENTER>.
Calibration of correction tables for Stereo is to be done
with Stereo angle 0
i
and with cassette in one of the end
positions of the object table.
1 Help 2 Save
Cal
3
4
5
6
7
8
9
10
Quit
M
A
M
0
0
7
3
1
<ESC> to exit, <F2> to save, <F3> to calibrate, <F4> to clear calibration
Object table group
Magnification
Not calibrated
Grid
Not
calibrated
No grid
Not calibrated
Stereo
Not calibrated
Mo/Mo H, Table ID: 1104
Clear
Fig. 4 Dialog
Calibration AEC correction table
shows which object tables, in the current Anode/Filter/Speed
Please mount a
grid table
and press <ENTER> to continue
or <ESC> to cancel.
<ENTER> - ok / <ESC> - cancel
M
A
M
0
0
6
0
8
NOTE
Used: 0
New: 0
<F2> save, <F3> exp support, <F4> continue, <ESC> exit
M
A
M
0
0
7
6
4
Sensitivity correction
Fig. 5 Dialog
Sensitivity correction
Calibration AEC correction table