M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 6 of 14
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
5
-
6
Cable
connections
Connecting the cable harness to the generator
5
Unit control cable X1
5
1. Connect the unit control cable X1 to D700 X1 (1).
Filament cable X8
5
1. Connect the filament cable X8 to D700 X8 (2).
Power supply cable X14
5
1. Connect the power supply cable X14 to D711 X14 (3).
2. Strain-relieve it on the side with a cable tie (4).
AEC signal cable X10 and X11
5
1. Connect the AEC signal cable connectors, X10 and X11, to D700 X10 (5) and X11
(6).
EMC measures required.
They are described on page 5 - 1 in this chapter.
Fig. 6 Cable harness at the generator
The AEC signal cable consists of two cables laid together as one
cable with two connectors labeled X10 and X11 at the end of the
cable.