Calibrating and adjusting the AEC
11 - 11
Siemens AG
Register 3
SPB7-230.033.11
Page 11 of 22
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
AEC Correction tables
11
This section will guide you through the following steps:
•
Installation of correction tables (F2).
•
Calibration of correction tables (F3), including rough setting of sensitivity correction.
The AEC uses one correction table for each Anode/Filter/Speed combination.
Each correction table has separate corrections for each object table group (Magnification,
Grid, No grid and Stereo).
DLF switch off
11
Deactivate the DLF switch. In <Main menu> select <Configuration>, <Miscellaneous> and
<DLF switch>, press space to set switch to OFF. Press <F2> to save.
Increase grid speed
11
Increase the grid speed. In <Main menu> select <Configuration> and <Grid speed>.
Set the following values:
For exact positioning of the needle, the anode/filter combination
"Mo/Mo" or "Mo/Rh" must be used. Therefore, calibration does not
have to be attempted for W/Rh combination in stereo mode.
NOTICE
Grid
1 Help
2 Save
3
4
5
6
7
8
9
10 Quit
Grid fast speed time
1500
Grid fast speed
90
Grid slow speed
30
ms (2.5 s max)
% of max
% of max
<ESC> to exit, <TAB> move to next entry field