Appendix
22
-
11
Siemens AG
Register 3
SPB7-230.033.11
Page 11 of 12
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Keep the measuring protocol with the technical documentation.
Tube P40 MoW, MAMMOMAT 3000 Nova,
tungsten anode 0.3, AEC-mode
40 mm Plexiglas
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5%
______V x 40 mA/V = ____ mA
158 mA
&
10%
Date:______________
Sign:____________
Tube P40 MoW, MAMMOMAT 1000/3000 Nova,
molybdenum anode 0.3, AEC-mode
40 mm Plexiglas
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5%
______V x 40 mA/V = ____ mA
125 mA
&
10%
Date:______________
Sign:____________