Checks
with
high
voltage
10
-
9
Siemens AG
Register 3
SPB7-230.033.11
Page 9 of 10
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
*
) total exposure time dependent on programmed sensitivity
Measurements in AEC mode shall be performed after calibrating
and adjusting the AEC.
Fig.
11
Fig.
12
30 kV
1 V/T (1 T <=> 5 kV)
30 kV
1 V/T (1 T <=> 5 kV)
158 mA
1 V/T (1 T <=> 40 mA)
125 mA
1 V/T (1 T <=> 40 A)
*
)
100 ms/T)
*
)
100 ms/T
NOTE
Tube P40 MoW,
MAMMOMAT 3000 Nova
* tungsten anode 0.3
Tube P40 MoW,
MAMMOMAT 1000/3000 Nova
molybdenum anode 0.3
M
A
M
0
0
6
6
6
30 kV AEC
(P = 4.7 kW)
40 mm plexiglass
M
A
M
0
0
6
6
7
30 kV AEC
(P = 3.75 kW)
40 mm plexiglass