Cable
connections
5
-
7
Siemens AG
Register 3
SPB7-230.033.11
Page 7 of 14
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Rotating anode cable X9
5
1. Connect the rotating anode cable X9 to D711 X9 (1).
2. Connect the two blue free wires from connector D711 X9 (1) to capacitor C51 (2).
3. Clamp the cable to the chassis with a cable clamp (3).
Fig. 7 Connecting X9 to the generator
Obtain proper grounding.
When fitting the cable clamp, ensure satisfactory contact
between shield braid (pos. 4 / Fig. 7) and chassis.