13 - 1
Siemens AG
Register 3
SPB7-230.033.11
Page 1 of 4
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Testing
13
Testing the AEC-function
13
AEC performance test
13
Testing of the AEC-function is to be performed according to the test protocol, and for the
object tables in question. (Should error codes Er013 or Er450 appear; change to AEC cal-
ibration plexi which is 1 cm thinner.) Mark the exposed films with exposure conditions and
store them together with the test protocol.
These limits apply for film with max. incremental gamma 5.0.
Tolerance only valid for film/screen combinations listed in
Table 1 under "Checking and programming with the service PC"
on Page 11 - 1.
It is important to make sensitometer exposures on all films to be
measured.
NOTE
NOTE