Oscilloscope diagrams
Tube P40 MoW, MAMMOMAT 3000 Nova, tungsten anode 0.3
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5%
______V x ____ s x 40 mA/V = ____ mAs
20 mAs
&
10%
Date:______________
Sign:______________
Tube P40 MoW, MAMMOMAT 1000/3000 Nova,
molybdenum anode 0.3
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5%
______V x ____ s x 40 mA/V = ____ mAs
20 mAs
&
10%
Date:______________
Sign:______________
Tube P40 MoW, MAMMOMAT 3000 Nova, tungsten anode 0.15
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5%
______V x ____ s x 40 mA/V = ____ mAs
10 mAs
&
10%
Date:______________
Sign:______________
Tube P40 MoW, MAMMOMAT 1000/3000 Nova,
molybdenum anode 0.15
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5%
______V x ____ s x 40 mA/V = ____ mAs
10 mAs
&
10%
Date:______________
Sign:____________