Preparatory
work
3
-
7
Siemens AG
Register 3
SPB7-230.033.11
Page 7 of 8
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Optional:
5. Insert sheet-metal strip (5/ Fig. 10) (additional radiation protection) into both
stanchions.
6. Fit the lead-glass (4/ Fig. 10) into both vibration absorbing strips (6/ Fig. 10) and fix
it softly by screws (1/ Fig. 10).
7. Fit the lead-glass (2/ Fig. 10) into the groove of one of the stanchions (3 / Fig. 10)
and keep it in position while mounting the other stanchion (3 / Fig. 10).
8. Mount the stanchions and lead-glass unit with 6 screws onto the generator
(3, 4 / Fig. 9) (two persons are required).
9. Cover the holes with plastic caps (7/ Fig. 10).
The screws must be mounted before the lead-glass is inserted in
the stanchions.
Fig. 10 Mounting the radiation shield
NOTE