M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 2 of 10
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
20
-
2
Final
procedures
Removing the measuring instruments
20
1. Turn OFF the generator.
2. Remove the service PC.
3. Remove connected measuring instruments.
Checking the protective grounding resistance
20
Using the protective ground-wire tester, measure the resistance between the protective
ground terminal in the stand (X899, see Chapter 5 "Cable connections") and all metallic
parts of the equipment which the patient/operator may come in contact with, incl. all object
tables. The protective ground resistance must not exceed 0.1
Ω
.
Mounting the cable duct covers
20
If not already done, mount the cable duct covers, see Chapter 4 "Installing the generator
and the stand".
Mounting the stand covers
20
The edges of the metal curtain of the stand are very sharp.
They may cause severe injury.
Remove the protective strips carefully when the covers are to be
mounted. Store the strips in the holders provided on both sides of
the curtain.
When mounting the stand covers, make sure that the screws at
the top of the stand are securely tightened and have functioning
contact washers.
These screws are used to establish protective ground connec-
tions.
WARNING
NOTICE