M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 2 of 4
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
13
-
2
Testing
Testing and adjusting OPDOSE
13
The unit is factory-set with the following values, which are to be regarded as recommen-
dations. Changes can be made with the service PC program (
Configuration, Miscella-
neous, Auto limits)
.
Thickness
kV
Anode/filter
combination
≥
61
mm
26
W
/
Rh
46
-
60
mm
27
Mo
/
Rh
31
-
45
mm
27
Mo
/
Mo
0
-
30
mm
26
Mo
/
Mo
Fig. 1 Installations with filter disc Mo 0.030 / Rh 0.025 /Rh 0.050
Thickness
kV
Anode/filter
combination
≥
61
mm
28
Mo
/
Rh
46
-
60
mm
27
Mo
/
Rh
31
-
45
mm
27
Mo
/
Mo
0
-
30
mm
26
Mo
/
Mo
Fig. 2 Installations with filter disc Mo 0.030 / Rh 0.025
Thickness
kV
Anode/filter
combination
≥
61
mm
29
Mo
/
Mo
46
-
60
mm
28
Mo
/
Mo
31
-
45
mm
27
Mo
/
Mo
0
-
30
mm
26
Mo
/
Mo
Fig. 3 MAMMOMAT 1000 installations with single filter from serial no. 7400 / software version 4.2