14 - 1
Siemens AG
Register 3
SPB7-230.033.11
Page 1 of 2
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Further programming
14
Setting the real time clock
14
Prerequisite
14
The correct date and time must be set in the service PC. If not set date at C:\> by typing
date
<
ENTER
>
and time at C:\> by typing
time <
ENTER
>.
Setting the time in the MAMMOMAT
14
Main menu:
1. Select <Configuration>.
2. Select <Clock>.
3. Press <F2>.
With <F2> the date and time set in the service PC are taken over by the
MAMMOMAT.
Reducing the generator power
14
To achieve full power, the measured resistance must not exceed the following values:
•
0.25
O
at 110 V (1-phase)
•
0.45
O
at 208 V
•
0.50
O
at 230 V
•
0.60
O
at 240 V
•
0.65
O
at 277 V
•
0.85
O
at 400 V (2-phase)
If the above values are exceeded, reduce the generator power as follows:
Main menu:
14
1. Select <Configuration> <Miscellaneous >.
2. Select <Red. of power>.
Reduction of power depending on line quality.
Tube is P 40 MoW.
3. Select nominal line voltage ____ V with cursor keys <
↑
> <
↓
> and < ENTER>.
4. Type in line impedance ____
Ω
. Enter the measured line impedance.
Calculated power is ___ kW. Max. possible tube power is displayed.
5. Save the programmed values with <F2>.
If the power is reduced after calibrating and adjusting the AEC, a
new calibration will be necessary!
NOTE