Calibrating and adjusting the AEC
11 - 5
Siemens AG
Register 3
SPB7-230.033.11
Page 5 of 22
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Error message
If the menu "Dos critical error" would appear during the installation, always press "F"
(for fail) and check that the correct floppy is inserted, and is not write protected.
6. Press <F3> and follow the instructions in the dialogs for exposure.
7. The measured dose rate will be displayed on the screen. If the result is OK go on
with step 8. If not OK, check that the detector is not mechanically grounded and
repeat the test.
8. Make two more exposures with the same Anode/Filter combination, by repeating
steps 6 and 7.
9. Repeat steps 6 to 8 until all exposure alternatives, as listed above, are covered.
10. Remove metal plate (part no. 65 52 728).
11. Exit the offset compensation test in service PC by pressing Esc.
If Er 02X occurs:
Exit the offset compensation test in the Service Program by
pressing Esc. Acknowledge the error by pressing the limit button.
Wait until the green lamp on the control panel is lit. In <Mainmenu>
select <Service> and <Offset compensation test> and continue the
test.
Make sure that the correct anode/filter combination and wing are
selected, according to the list of alternatives at the top of this
page.
M
A
M
0
0
7
3
5
NOTICE
NOTE
[DOS Critical Error]
[
]