Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Attaching the swivel-arm covers
7
Attaching the swivel-arm covers
7
Arranging the swivel-arm system
7
1. Install object table, no compression force (swivel-arm system still upside down).
2. Remove the protective strips from the metal curtain.
3. Run the lifting carriage upwards.
4. Put back the protective strips onto the metal curtain.
5. Rotate the swivel-arm system to 0 degrees by pressing one of the switches for
clockwise rotation on the supporting arm.
6. System and mains "OFF".
7. Remove the swivel-arm system transport safeguard, see Chapter 3
"Preparatory work".
Connecting the cables to control-button boards and patient handles
7
Before installing the side covers, the cables are to be connected.
1. Connect cables X807 (1) to control-button circuit boards D807 (2).
2. Connect ground wires (3) to patient handles (4).
Fig. 1 Connecting the cables
M
A
M
0
0
1
0
8
2
1
4
3
2
1
4