Calibrating and adjusting the AEC
11 - 21
Siemens AG
Register 3
SPB7-230.033.11
Page 21 of 22
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Sensitivity
11
Adjustment of O.D. for the whole AEC is performed under <Mainmenu>, <Configuration>,
<AEC> and <Sensitivity>.
The O.D. preferred by the customer can be set with this
function. The customer’s own phantom(s) can also be
used.
1
Help
2
3
5
6
7
8
9
10
Quit
<ESC> to exit, <TAB> move to next entry field
M
A
M
0
0
6
5
4
H
D
valid entries from -64 to 64 in 1/8 E.P. (+/-8 E.P.)
Save
Sensitivity
0
0
NOTE
AEC Sensitivity