M1000/3000 Nova
Register 3
SPB7-230.033.11
Page 2 of 2
Siemens AG
Installation and Start-Up
Rev. 06
05.06
CS PS SP
Medical Solutions
16
-
2
Field
light
Adjusting the field light time
16
The illumination time of the light field can be adjusted according to the customer´s desire
within a certain range.
1. Connect the service PC to the generator and start the program as described in
Chapter 22 "Appendix".
2. Program according to the following:
Main menu:
1. Select <Configuration>.
2. Select <Miscellaneous>.
3. Select <Illumination time>.
4. Type in illumination time and save with < F2>.
Shorter illumination time means increased life time of the lamp.
NOTE